@Article{, title={Quality assurance of the linear accelerator device using star track and Perspex ضمان جودة جهاز المعجل الخطي باستخدام ستار تراك وبيرسبكس}, author={Basma Saber Lazim بسمة صبر لازم and Basim Khalaf Rejah باسم خلف رجه and Haydar H. Alabdi حيدر حمزة العبادي}, journal={Iraqi Journal of Physics المجلة العراقية للفيزياء}, volume={17}, number={42}, pages={51-55}, year={2019}, abstract={Radiation therapies with linear accelerator are one of the efficient techniques cancer treatment. The aim of this study is to verify of quality assurance of the linear accelerator at the Baghdad Center for Radiation Therapy and Nuclear Medicine. In this study, a STARTRAK device and Perspex were used to ensure the quality of the linear accelerator for the period from August to December 2018. Quality assurance of the linear accelerator work using Perspex and Star Track by measuring the output dose. The results show an acceptable variation in the output X-ray dose delivered from Linear accelerator type synergy. The achieved variation was ±2% and it was within the permissible range according to the recommendations of the manufacturer of the accelerator (Elkta).

تعد العلاجات الإشعاعية باستخدام المعجل الخطي واحدة من الأساليب الفعالة لعلاج السرطان. حيث كان الهدف من هذه الدراسة هو التحقق من ضمان جودة المعجل الخطي المتوفر في مركز بغداد للعلاج الإشعاعي والطب النووي باستخدام ستار تراك وبيرسبكس. اجريت الدراسة في الفترة من شهر اب إلى شهر كانون الاول 2018. أظهرت النتائج تباينًا مقبولًا في الأشعة السينية الخارجه من المعجل الخطي السنرجي. هذا الاختلاف بلغ ± 2٪ وهو ضمن النطاق المسموح به وفقًا لتوصيات الشركة المصنعة للمعجل (اليكتا).} }