@Article{, title={Evaluation of Sysmex Automated Hematological Analyzer (KX-21N) for the measurement of Blood Hemoglobin Level}, author={Yesar M.H.Al-Shamma MBChB.,PhD.(UK).}, journal={KUFA MEDICAL JOURNAL مجلة الكوفة الطبية}, volume={15}, number={1}, pages={1-5}, year={2012}, abstract={Abstract: This study was carried out on twenty normal subjects (women) in order to evaluate (Sysmex Automated Hematological Analyzer KX-21N) for the estimation of blood hemoglobin level. Two processes where used for evaluation, the first is the systematic error to determine the accuracy of the device by comparison with the standard device (the Sahli's) and the second process is to test the random error or reproducibility of Sysmex. The result of this study indicate that values of Hb measured by Sysmex KX-21N were slightly higher than that measured by Sahli's method (P<0.005) And the result of random error test indicates that the Sysmex KX-21N is reproducible device which is improved by using the statically equation: (2SD/mean x 100%) to determine 95% tolerance limit of the device which indicate that the device is greatly reproducible.

أجريت هذه الدراسة على عشرين شخصا من الأصحاء لغرض تقييم جهاز فحص نسبة الهيموغلوبين في الدم Sysmex KX-21N)) وذلك بإتباع طريقتين: أولا مقارنته بطريقة Sahli)) والتي تعتبر الطريقة القياسية في قياس مستوى الهيموغلوبين لمعرفة مدى دقة الجهاز المراد تقييمه. ثانيا لمعرفة دقة الجهاز أثناء القياسات المتكررة و التي أظهرت نتائجها بان القياسات المتتالية للجهاز متقاربة جدا في القيمة و لغرض البرهنة على دقة الجهاز استعملت المعادلة الإحصائية: (2SD/mean x 100%)والتي تؤكد على أن جهاز يعطي قراءات دقيقة ويمكن استعماله في البحوث الطبية والمستشفيات الغرض منه قياس مستوى الهيموغلوبين في الدم عند الأصحاء وعند المرضى.} }