@Article{, title={A study of the response of solid state nuclear track detectors LR115-Ι, LR115-П, CN85 and CR39 to alpha particles using track diameter measurement دراسـة استجابة كواشف الآثار النووية الصلبة, LR115-П, LR115-Ι CN85 و CR39 لجسيمات آلفا}, author={H.Bakr د.حمــزة بكــر سلمــان}, journal={Journal of Basrah Researches (Sciences) مجلة ابحاث البصرة ( العلميات)}, volume={33}, number={2A}, pages={83-90}, year={2007}, abstract={The accuracy of track diameter for (LR115-Ι, LR115-П, CN85 and CR39) are measured in the present work. Comparison for the response curve is studied regarding the relation between the growth of the diameter (Vd ) and the energy of alpha particles for different detectors with the given processing parameters. As a conclusion: alpha particles with energy 4 MeV give the better results.CR39 track detector is more effected by the energy of alpha particles and the response to register the charge particles than the other detectors used.

تم في البحث الحالي دراسة الدقة في قطر الأثر للكواشف (, LR115-П, LR115-Ι CN85 و CR39) . ودرس منحني الاستجابة من خلال العلاقة بين نمو أقطار الآثار (Vd ) وطاقة جسيمات آلفا للكواشف المستخدمة وبشروط معينة. استنتج من الدراسة الحالية بان طاقة (4 MeV) لجسيمات آلفا تعطي افضل نتائج . وأن كاشف الأثر الصلب CR39 هو أكثر الكواشف استجابة لجسيمات آلفا من الكواشف الأخرى المستخدمة في هذا البحث.} }