Preparation of New Copper (II) – PVC Membrane Sensor Based on Phenyl Disulfide

Aayad Ammar Sayhood

journal of kerbala university مجلة جامعة كربلاء
ISSN: 18130410 Year: 2011 Volume: 9 Issue: 3 Pages: 93-101
Publisher: Kerbala University جامعة كربلاء

Abstract

Membrane of phenyl disulfide as an ionic membrane carrier in poly (vinyl chloride) (PVC) based matrix have been used for preparing Copper selective electrode. The electrode exhibits according to Nernst equation response for Cu2+ ions over a wide concentration range from 1.0x10-6 to 1.0x10-1 M with a slope equal to 29.2 mV/decade, r=0.9988 and a detection of limit 4.4x10-7 M. The response time measured was 8 second. It was found to be selective and useable within the pH range of (2.0-7.0) and the temperature equal to 293 K to 303 K. The lifetime of membrane sensor prepared could be used for at least 3 months. The electrode was successfully used as an indicator electrode in potentiometric titration of Copper ions with EDTA.

غشـاء الناقـل الايـوني phenyl disulfide في قالب البولي (فنيل كلورايد) (PVC) استخدم لتحضير قطب النحاس الانتقائي . أظهر القطب استجـابة حسب معادلة نيرنست لايونــات Cu2+ ضمن مدى تراكيز تراوحت بين (1.0x10-1 -1.0x10-6) مولاري وبميـل مقداره 29.2 decademV وقيمة r تساوي 0.9988 وحـــد تحسس مـقداره4.4x10-7 مولاري . وكان زمـن الاستجابة 8 ثواني . تم إيجاد الانتقائية ومـدى الدالة الحامضية المناسبة والتي تراوحت ما بين (7.0-2.0) ودرجة الحرارة مساوية الى (293-303) كلفن. وكان العمر الزمني للغشاء المحضر ثلاثة اشهر . وتم استعمال القطب بنجاح كقطب دليل في التسحيح الجهدي لايونات النحاس مع EDTA .

Keywords

Copper selective electrode --- membrane --- sensor --- response time --- lifetime --- potentiometric titration.